IPM300 Series: Analog Output
 

The analog output function will allow you to output a signal to another source. The strength of this signal is based on the input signal from the load on the sensor.
Wiring Directions..

Note: Display will automatically exit setup mode and return to run mode if a button is not pushed within 30 seconds.

1. Press both the buttons at once, ‘CAL’ will appear, press the bottom button until ‘AnOU’ appears. Press the top button.
2. ‘ruG’ will appear, press the top button to change the range of the analog output signal. The range is up to the user. NOTE: the first two setting ‘0-20’ and ‘4-20’ are currents and the last three are voltages. Once the display is showing your chosen range press the top button. ‘rnG’ will then appear. Press the bottom button.
3. ‘StLo’ will appear. Press the top button. Here you are to enter the display reading that will correspond with the low output signal (0000). Press the top button to save your value and ‘StLo’ will appear again. Press the bottom button.
4. ‘StHi’ will appear. Press the top button. Here you are to enter the display reading that will correspond with the high output signal. Press the top button to save your value and ‘StHi’ will appear again. Press the bottom button.
5. ‘SOUr’ will appear. Press the top button. Here you have an option of the source of data driving the output signal. The options are listed below:
‘rEAd’ – display reading
‘ nEt’ – filtered net value
‘GrOS’- filtered gross value
‘ntFS’ – unfiltered net value
‘GrFS’ – unfiltered gross value
‘PK 1’ – Peak 1 stored value
‘PK 2’ – Peak 2 stored value
Select your desired source by pushing the
bottom button to scroll down the menu and the top button when your choice is displayed on the meter. ‘SOUr’ will appear again, press the bottom button.
6. ‘End’ will now appear. Press the top button to exit the menu or the bottom button to reset the analog output.
 
Support
 
Web Feature
 
Shipment Tracker
Use your order number and e-mail to track your shipment!
Email:
Order #:
 
Live Support
Live Support
Need live help from our web and sales team? Click here.
 
Events
No Current Events
 
Newsletter
Newsletter
want to stay informed?
Sign up for our Monthly Newsletter!
Credit CardsISO 17025:2005 A2LA Accredited Calibrations
US ManufacturerUS Manufacturer