IPM300 Series: Automatic Calibration
 

This function allows you to calibrate the unit to an input signal. This method works by applying a low input signal and a corresponding display value and a high input signal and a corresponding display value. One benefit of this method is that the individual signals can be applied at any given time. It is possible to input the low signal at one time and the high signal at a completely different time. This calibration option is located under the ‘SPEC’ menu.

Note: Display will automatically exit setup mode and return to run mode if a button is not pushed within 30 seconds.

1. Press both buttons and ‘CAL’ will appear. Enter the ‘CAL’ menu by pressing the top button. Then press the bottom button until ‘rdLO’ appears. Press the top button and enter in a low display value that will correspond to your low input signal. Press the top button until the meter reads ‘rdLO’. Press the bottom button.
2. ‘rdHI’ will now be on the display. Press the top button and enter the high value that should be displayed when the high input signal is applied. Press the top button until the display reads ‘rdHI’.
3. Exit the setup mode and reset the power to the meter.
4. Press both buttons, ‘CAL’ will appear, press the bottom button until ‘SPEC’ appears. Press the top button to enter the special menu. Press the bottom button until ‘ACAL’ appears. Press the top button.
5. ‘ACLO’ will appear. Have your minimum load on the sensor and press the top button. The meter will flash ‘CAH 0’ and the reading from the load. Press the bottom button. The meter will then exit out of the setup mode and display the input signals.
6. If you are ready to enter in the high signal you can do it now.
7. Press both buttons and enter the main menu. Return to the ACAL menu (directions above) and push the bottom button until ‘ACHI’ appears. Apply your load to the sensor and press the top button. The display will then flash ‘CAH 1’ and the input signal. Press the bottom button, the value will be saved and the meter will return to the run mode.
8. Once again enter the ACAL menu, press the bottom button until ‘CALC’ appears. Press the top button and the meter will activate the calculations and go through the reset procedure. The unit is now ready to use.
 
Support
 
Web Feature
 
Shipment Tracker
Use your order number and e-mail to track your shipment!
Email:
Order #:
 
Live Support
Live Support
Need live help from our web and sales team? Click here.
 
Events
No Current Events
 
Newsletter
Newsletter
want to stay informed?
Sign up for our Monthly Newsletter!
Credit CardsISO 17025:2005 A2LA Accredited Calibrations
US ManufacturerUS Manufacturer