To Set the Inflection Point Setting: If
you select the inflection point hold in the hold functions settings, then it is necessary to set the inflection point
parameters, the ‘Minimum Slope’, ‘Interval A’ and ‘Interval B’.
- Press the Mode button
- Select the 'HOLD' tag
- Press the Page button until Minimum Slope appears. Select the Minimum Slope tag.
- Enter in the minimum slope with the numerical keys, then press OK.
- Press the Interval A tag and enter your desired interval (010-120), press OK.
- Press the Interval B tag and enter your desired interval (010-120), press OK
- Press the HOME button to return to the run mode
|
|