There are two excitation values available on the CSG110, 10 VDC (default) and
5 VDC. To select your excitation, simply connect the corresponding jumper
The input value range is from 0.5 to 4.0 mV/V (based on 10VDC excitation).
Select the input value closest to your input range by soldering the
corresponding jumper
For example: If you
are using a 2mV/V sensor with the 5 VDC excitation, then you would want to
solder the 1.0 (10mV) jumper. If you are using a 2mV/V sensor with 10 VDC
excitation then you would want to solder the 2.0 (20mV) jumper.